Diffusion Layer defines the interrogated Volume

Bard discusses the diffusion layer in the first chapter, section 1.4 (introduction to mass transfer controlled reactions). This 1D diffusion layer encompasses the linear concentration gradient. Beyond this distance, the concentration is effectively the same as the bulk solution.

\[\delta(t)=\text{ diffusion layer over time}\]

The moles of a molecule electrolyzed (M) at an electrode is approximately the difference in concentration between the bulk solution and electrode multiplied by the volume encompassed by the area of the electrode and half the diffusion layer. \[M \approx [C_{bulk}-C_{electrode}] \frac{A\delta(t)}{2}\]

For an electrode potential where a molecule is completely reduced/oxidized the \(C_{electrode}\) is effectively zero. \[\text{Where }C_{electrode}=0 \]

\[M = C_{bulk} \frac{A\delta(t)}{2}\]

This equation should essentially be: \[ Mass = Concentration * Volume \]

Therefore: \[ Volume = A*\frac{\delta(t)}{2}\]

Since volume can be described by an area times height (z): \[ Volume = A*z \]

\[ z = \frac{\delta(t_s)}{2}\]

The diffusion layer is defined in terms of \(D_{ap}\) and the scan time \(t_s\): \[\delta(t) = 2 \sqrt{D_{ap}t_s}\]

So finally z can be defined purely in terms of \(D_{ap}\) and the scan time \(t_s\): \[ z = \sqrt{D_{ap}t_s}\]

Therefore the volume A*z interrogated by the electrode depends on Dap (and scan time)!

Math for \(D_{phys}\) estimation (depends on volume/z)

Recall that for \(D_{phys}\) We start with the dilution of an initial mass with 1D diffusion next to a no flux boundary: \[C(t) = \frac{2M_0}{A_{xy}\sqrt{4\pi D_{phys}t}} \tag{eq. 1}\] Peak Square Wave current is given by: \[I_{swv} = \frac{nFA_{xy}C \sqrt{D_{ap}}}{\sqrt{\pi t_p}} \psi \tag{eq. 2}\] Concentration can be broken into mass / volume \[C = \frac{M}{V} = \frac{I_{swv}\sqrt{\pi t_p}}{nFA_{xy}\sqrt{D_{ap}}\psi} \tag{eq. 3}\] Therefore initial mass would be: \[M_0 = \frac{I_0\sqrt{\pi t_p}V}{nFA_{xy}\sqrt{D_{ap}}\psi} \tag{eq. 4}\] Substituting into eq. 1 gives: \[\frac{I_{swv}\sqrt{\pi t_p}}{nFA_{xy}\sqrt{D_{ap}}\psi} = \frac{2 \frac{I_0\sqrt{\pi t_p}V}{nFA_{xy}\sqrt{D_{ap}}\psi}}{A_{xy}\sqrt{4\pi D_{phys}t}} \tag{eq. 5}\] Simplified: \[I_{swv} = \frac{2I_0V}{A_{xy} \sqrt{4 \pi D_{phys}t}} \tag{eq. 6}\] The relevant volume should be the surface area of the electrode * the height of the biofilm? \[\frac{V}{A_{xy}}=z \tag{eq. 7}\] Below I plot this normalized decay curve: \[\frac{I_{swv}}{I_0} = \frac{z}{\sqrt{ \pi D_{phys}}}t^{-1/2} \tag{eq. 8}\] Fitting the normalized decay curve againt t^-(1/2) yields this slope: \[slope = m = \frac{z}{\sqrt{ \pi D_{phys}}} \tag{eq. 9}\] Rearranging gives Dphys: \[D_{phys} = \frac{z^2}{\pi m^2} \tag{eq. 10}\]

Math for \(D_{ap}\) estimation from titration (does not depend on volume/z)

Slope of the linear fit of GC vs. SWV titration plot: \[slope = m = \frac{\Delta I_{gc}}{\Delta I_{swv}} \tag{eq. 1}\] Generator collector peak current: \[I_{gc} = nFSCD_{ap} \tag{eq. 2}\]
Square Wave peak current: \[I_{swv} = \frac{nFACD_{ap}^{1/2}}{\pi^{1/2}t_p^{1/2}} \psi \tag{eq. 3}\] Substituting into eq. 1: \[m = \frac{I_{gc}}{I_{swv}} = \frac{nFSCD_{ap}}{\frac{nFACD_{ap}^{1/2}}{\pi^{1/2}t_p^{1/2}} \psi} \tag{eq. 4}\] Simplifying: \[m = \frac{S\pi^{1/2}t_p^{1/2}D_{ap}^{1/2}}{A\psi} \tag{eq. 5}\]

\[m^2 = ({\frac{S\pi^{1/2}t_p^{1/2}D_{ap}^{1/2}}{A\psi}})^2 \tag{eq. 6}\]

\[m^2 = \frac{S^2\pi t_p D_{ap}}{A^2\psi^2} \tag{eq. 7}\] Rearranged for Dap: \[D_{ap} = \frac{m^2 A^2 \psi^2}{S^2\pi t_p} \tag{eq. 8}\]